FDA Adverse Event Malfunction Summary report: N

PLMA DVC V11.51 1 N

MDR report key: 3061248 · Received April 10, 2013

Report

Report Number
9615050-2013-00655
Event Type
Malfunction
Date Received
April 10, 2013
Date of Event
January 1, 2013
Report Date
March 14, 2013
Manufacturer
HOSPIRA COSTA RICA LTD.
Product Code
FRN
PMA / PMN Number
K070398
Removal / Correction Number
Z-0624-2013
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

A FIELD SERVICE ENGINEER PERFORMED TESTING AND INVESTIGATION AT THE USER FACILITY. DURING TESTING, IT WAS NOTED THAT THE DOOR ROLLER WAS BROKEN. THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.

Description of Event or Problem · 1

THE CUSTOMER CONTACT REPORTED THAT DURING PREVENTIVE MAINTENANCE TESTING AT THE USER FACILITY, IT WAS NOTED THAT THE DEVICE ROLLER WAS BROKEN. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS WHILE THE DEVICE WAS IN CLINICAL USE. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
153635 PLMA DVC V11.51 1 N 80FRN FRN HOSPIRA COSTA RICA LTD. NA NA

Patients

Seq Age Sex Outcome Treatment
1 NA PLUM A+ SOFTWARE MODULE: LIST # 12097, SN UNK