PLUM A + DRIVER NEW
Report
- Report Number
- 9615050-2013-00656
- Event Type
- Malfunction
- Date Received
- April 10, 2013
- Date of Event
- January 1, 2013
- Report Date
- March 15, 2013
- Manufacturer
- HOSPIRA COSTA RICA LTD.
- Product Code
- FRN
- PMA / PMN Number
- K042081
- Removal / Correction Number
- Z-0624-2013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
A FIELD SERVICE ENGINEER PERFORMED TESTING AND INVESTIGATION AT THE USER FACILITY. DURING TESTING, IT WAS NOTED THAT THE DEVICE DOOR ROLLER WAS MISSING AND THE ROLLER PIN WAS BROKEN. THE DEVICE HAS BEEN IDENTIFIED AS PART OF A PRODUCT RECALL. THIS REPORT REPRESENTS ALL THE INFORMATION KNOWN BY THE REPORTER UPON QUERY BY HOSPIRA PERSONNEL.
THE CUSTOMER CONTACT REPORTED THE DEVICE DOOR ROLLER WAS MISSING AND THE ROLLER PIN WAS BROKEN. THE DEVICE WAS RETURNED TO THE BIOMEDICAL DEPARTMENT WITH AN UNSIGNED NOTE THAT STATED, "BATTERY NOT WORKING." NO TRACKING INFORMATION WAS PROVIDED; THEREFORE, SPECIFIC PATIENT INFORMATION, DEVICE PROGRAMMING, OR EVENT DETAILS WERE NOT AVAILABLE. THERE WERE NO REPORTS OF ANY ADVERSE PATIENT EVENTS OR DELAYS IN CRITICAL THERAPIES WHILE THE DEVICE WAS IN CLINICAL USE. DURING TESTING AT THE USER FACILITY, IT WAS NOTED THAT THE DEVICE DOOR ROLLER WAS MISSING AND THE ROLLER PIN WAS BROKEN. THOUGH REQUESTED, NO ADDITIONAL INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 148211 | PLUM A + DRIVER NEW | 80FRN | FRN | HOSPIRA COSTA RICA LTD. | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | UNK | SN (B)(4)| PLUM A+ SOFTWARE MODULE: LIST #20791 |