FDA Adverse Event Malfunction Summary report: N

TRILOGY 100

MDR report key: 3061045 · Received April 9, 2013

Report

Report Number
2518422-2013-00555
Event Type
Malfunction
Date Received
April 9, 2013
Date of Event
March 21, 2013
Report Date
March 21, 2013
Manufacturer
RESPIRONICS, INC.
Product Code
CBK
PMA / PMN Number
K083526
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DURING THE EVAL OF THE DEVICE AT THE MFR'S SERVICE CENTER, "SERVICE REQUIRED" CODES WERE FOUND IN THE VENTILATOR'S DOWNLOADED ERROR LOG. THE DEVICE'S SENSOR BOARD, POWER MANAGEMENT BOARD AND INTERNAL BATTERY WILL BE REPLACED TO ADDRESS THE ISSUES. DEVICE HAS BEEN EVALUATED, BUT THE COMPONENTS HAVE NOT BEEN REPLACED PENDING CUSTOMER APPROVAL OF ESTIMATE.

Description of Event or Problem · 1

A VENTILATOR WAS RETURNED TO THE MFR FOR SERVICE. THERE WAS NO HARM OR INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
148001 TRILOGY 100 VENTILATOR, CONTINUOUS, FACILITY USE CBK RESPIRONICS, INC. 1054096

Patients

Seq Age Sex Outcome Treatment
1