FDA Adverse Event Malfunction Summary report: N

TRILOGY 202

MDR report key: 3061043 · Received April 9, 2013

Report

Report Number
2518422-2013-00556
Event Type
Malfunction
Date Received
April 9, 2013
Date of Event
March 21, 2013
Report Date
March 21, 2013
Manufacturer
RESPIRONICS, INC.
Product Code
CBK
PMA / PMN Number
K093905
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DURING THE EVAL OF THE DEVICE AT THE MFR'S SERVICE CENTER, A "VENTILATOR INOPERATIVE" CODE WAS FOUND IN THE VENTILATOR'S DOWNLOADED ERROR LOG. THE DEVICE'S BLOWER MOTOR WAS REPLACED TO ADDRESS THE ISSUE. THE VENTILATOR FAILED A STEP DURING TESTING. THE DEVICE'S ACTIVE EXHALATION CONTROL MODULE WAS REPLACED TO ADDRESS THE ISSUE.

Description of Event or Problem · 1

A VENTILATOR WAS RETURNED TO THE MFR FOR ROUTINE PREVENTIVE MAINTENANCE. THERE WAS NO ALLEGATION OF DEVICE MALFUNCTION. THERE WAS NO HARM OR INJURY REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
144458 TRILOGY 202 VENTILATOR, CONTINUOUS, FACILITY USE CBK RESPIRONICS, INC. 1040007

Patients

Seq Age Sex Outcome Treatment
1