FDA Adverse Event Malfunction Summary report: N

TRILOGY 100

MDR report key: 3061042 · Received April 9, 2013

Report

Report Number
2518422-2013-00557
Event Type
Malfunction
Date Received
April 9, 2013
Date of Event
March 21, 2013
Report Date
March 21, 2013
Manufacturer
RESPIRONICS, INC.
Product Code
CBK
PMA / PMN Number
K083526
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DURING THE EVAL OF THE DEVICE AT THE MFR'S SERVICE CENTER, THE DEVICE FAILED A STEP DURING TESTING. THE DEVICE'S ACTIVE EXHALATION CONTROL MODULE WAS REPLACED TO ADDRESS THE ISSUE.

Description of Event or Problem · 1

A VENTILATOR WAS RETURNED TO THE MFR FOR ROUTINE PREVENTIVE MAINTENANCE. THERE WAS NO ALLEGATION OF DEVICE MALFUNCTION. THE DEVICE WAS NOT IN PT USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
148000 TRILOGY 100 VENTILATORY, CONTINUOUS, FACILITY USE CBK RESPIRONICS, INC. 1054655

Patients

Seq Age Sex Outcome Treatment
1