ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2013-04325
- Event Type
- Malfunction
- Date Received
- April 16, 2013
- Report Date
- March 19, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
FOLLOW-UP #1 DATE OF SUBMISSION (B)(4) 2013 - DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: INVESTIGATION REVEALED THAT THE DISPLAY SCREEN WAS DIM, FADED, AND DISCOLORED AND DIFFICULT TO READ. THE DISPLAY WAS REPLACED WITH A TEST DISPLAY, AND THE CONTRAST RETURNED TO NORMAL WITH NO SIGNS OF FADING OR DISCOLORATION.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. (B)(6).
ON (B)(6) 2013 THE REPORTER CONTACTED ANIMAS, ALLEGING THAT THE DISPLAY SCREEN WAS FADED/DISCOLORED. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE DEVICE CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.