FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3060948 · Received April 16, 2013

Report

Report Number
2032227-2013-01430
Event Type
Malfunction
Date Received
April 16, 2013
Date of Event
April 2, 2013
Report Date
April 2, 2013
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AL
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY, IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

THE INSULIN PUMP RECEIVED WITH A CRACKED RESERVOIR TUBE LIP. NO LOOSE DRIVE SUPPORT DISK NOTED.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT SHE HAD HIGH BLOOD SUGARS. CUSTOMER STATED THAT THE DRIVE SUPPORT CAP IS LOOSE ON HER INSULIN PUMP AND WHEN SHE PUSHED THE CAP IN MAKES A CLICKING NOISE AND THE INSULIN SQUIRTED OUT. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
162194 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-523LNAP

Patients

Seq Age Sex Outcome Treatment
1 48 YR