FDA Adverse Event Injury Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3060947 · Received April 16, 2013

Report

Report Number
2032227-2013-01429
Event Type
Injury
Date Received
April 16, 2013
Date of Event
March 26, 2013
Report Date
April 2, 2013
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CT
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

MOTOR ERROR ALARM DURING REWIND DUE TO CORRODED MOTOR HOME SWITCH. UNABLE TO PERFORM ALL FUNCTIONAL TESTING INCLUDING THE DISPLACEMENT, BASIC OCCLUSION, OCCLUSION, PRIME AND EXCESSIVE NO DELIVERY TEST DUE TO MOTOR ERROR ALARM.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT SHE HAD UNEXPLAINED HIGH BLOOD GLUCOSE. CUSTOMER WENT TO THE EMERGENCY ROOM AND WAS HOSPITALIZED. CUSTOMER'S BLOOD GLUCOSE READING AT THE TIME OF THE EMERGENCY ROOM VISIT WAS OVER 400 MG/DL. CUSTOMER STATED THAT HER INFUSION SET CANNULA WAS BENT. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
163854 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-522LNAL

Patients

Seq Age Sex Outcome Treatment
1 46 YR Hospitalization