PARADIGM REAL-TIME INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2013-01429
- Event Type
- Injury
- Date Received
- April 16, 2013
- Date of Event
- March 26, 2013
- Report Date
- April 2, 2013
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
MOTOR ERROR ALARM DURING REWIND DUE TO CORRODED MOTOR HOME SWITCH. UNABLE TO PERFORM ALL FUNCTIONAL TESTING INCLUDING THE DISPLACEMENT, BASIC OCCLUSION, OCCLUSION, PRIME AND EXCESSIVE NO DELIVERY TEST DUE TO MOTOR ERROR ALARM.
CUSTOMER REPORTED THAT SHE HAD UNEXPLAINED HIGH BLOOD GLUCOSE. CUSTOMER WENT TO THE EMERGENCY ROOM AND WAS HOSPITALIZED. CUSTOMER'S BLOOD GLUCOSE READING AT THE TIME OF THE EMERGENCY ROOM VISIT WAS OVER 400 MG/DL. CUSTOMER STATED THAT HER INFUSION SET CANNULA WAS BENT. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 163854 | PARADIGM REAL-TIME INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-522LNAL |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 46 YR | Hospitalization |