FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 3060871 · Received April 5, 2013

Report

Report Number
1720753-2013-04653
Event Type
Malfunction
Date Received
April 5, 2013
Date of Event
March 20, 2013
Report Date
April 5, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE POWER SUPPLY VOLTAGE WAS ADJUSTED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM DISPLAYED A PANEL COMMUNICATION ERROR MESSAGE. THIS ERROR MESSAGE IS LIKELY TO RESULT IN A SYSTEM LOCK UP, NO BOOT, OR SHUT DOWN. THERE IS NO REPORT OF PT INJURY ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
139777 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1