XMAX MOTOR

MDR report key: 3060791 · Received April 5, 2013

Report

Report Number
1045834-2013-01315
Event Type
Malfunction
Date Received
April 5, 2013
Report Date
April 6, 2012
Manufacturer
THE ANSPACH EFFORT, INC.
Product Code
ERL
PMA / PMN Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE HAS BEEN RECEIVED BY ANSPACH. RELIABILITY ENGINEERING EVALUATED THE DEVICE, AND THE REPORTED PROBLEM WAS CONFIRMED. THE MOTOR EXHIBITED DAMAGE TO THE LOCKING MECHANISM THAT CREATED FRICTION THAT PRODUCED THE REPORTED HEAT. THE DAMAGE TO THE MOTOR'S LOCKING MECHANISM WAS CONSISTENT WITH EITHER IMPROPER ASSEMBLY OR POWER BRAKING. IF ADDITIONAL INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SENT.

Description of Event or Problem · 1

REPORT RECEIVED FROM (B)(6) STATING THE DEVICE WAS "HEATING." THE DEVICE WAS BEING USED DURING A PROCEDURE. THE DATE OF THE VENT IS UNK. NO PT OR USER INJURIES WERE REPORTED. THERE WAS NO ADDITIONAL INFORMATION PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
141601 XMAX MOTOR ERL THE ANSPACH EFFORT, INC.

Patients

Seq Age Sex Outcome Treatment
1