FDA Adverse Event Malfunction Summary report: N

OLYMPUS THUNDERBEAT 5MM, 20CM, INLINE GRIP

MDR report key: 3060748 · Received March 29, 2013

Report

Report Number
8010047-2013-00080
Event Type
Malfunction
Date Received
March 29, 2013
Date of Event
March 4, 2013
Report Date
March 4, 2013
Manufacturer
OLYMPUS MEDICAL SYSTEMS CORPORATION
Product Code
KNS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE SUBJECT DEVICE WAS RETURNED TO OLYMPUS MEDICAL SYSTEMS CORPORATION (OMSC) FOR EVALUATION. THE EVALUATION CONFIRMED THAT THE PROBE BROKE DOWN AT 13 MM FROM THE DISTAL END. AND THERE WAS A SCRATCH AT THE BROKEN POINT. THERE WAS A CONTACT MARK OF THE PROBE AND JAW WHERE THE PROBE WAS BROKEN OFF. THE PTFE PAD WAS DEFORMED. THE MANUFACTURING HISTORY WAS REVIEWED, WITH NO IRREGULARITIES NOTED. BASED ON THE ANALYSIS RESULT ABOVE, SINCE THE DEVICE WAS GRASPING A THICK AND HARD TISSUE AND ACTIVATE WHILE TWISTING THE TISSUE, THE PTFE PAD UNEVENLY WORN AND THE PROBE AND JAW CAME INTO CONTACT WITH EACH OTHER SPARKS OCCURRED BETWEEN THEM: THAT CAUSED THE DEVICE A CRACK OCCURRED. THE TB-0520IC INSTRUCTION MANUAL ALREADY STATES; WARNING: DO NOT APPLY ONLY THE PROBE TIP TO THE TISSUE WITH A STRONG FORCE OR PULL THE PROBE TIP UP GRASPING A TISSUE OR ACTIVATE OUTPUT WHILE TWISTING THE SHAFT OR TURNING THE ROTATION KNOB WITH THE TRANSDUCER AND CONNECTION CABLE. OTHERWISE, THE PROBE MAY BE DAMAGED BY INTERFERENCE WITH THE INTERNAL PARTS, WHICH COULD RESULT IN THE TIP BREAKING AND DROPPING INSIDE THE BODY CAVITY. THIS REPORT IS BEING SUBMITTED AS A MEDICAL DEVICE REPORT IN AN ABUNDANCE OF CAUTION.

Description of Event or Problem · 1

WHEN A PHYSICIAN USED THE SUBJECT DEVICE FOR A PROCEDURE, THE PROBE OF THE SUBJECT DEVICE BROKE. THE BROKEN PIECE OF THE PROBE WAS TAKEN OUT. THE PHYSICIAN REPLACED THE SUBJECT DEVICE WITH A SIMILAR DEVICE. THE PROCEDURE WAS COMPLETED AS SCHEDULED. THERE WAS NO PATIENT HARM REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
129741 OLYMPUS THUNDERBEAT 5MM, 20CM, INLINE GRIP THUNDERBEAT HANDPIECE KNS OLYMPUS MEDICAL SYSTEMS CORPORATION TB-0520IC 29K

Patients

Seq Age Sex Outcome Treatment
1