OLYMPUS THUNDERBEAT 5MM, 45CM, PISTOL GRIP
Report
- Report Number
- 8010047-2013-00060
- Event Type
- Malfunction
- Date Received
- March 29, 2013
- Date of Event
- February 27, 2013
- Report Date
- March 4, 2013
- Manufacturer
- OLYMPUS MEDICAL SYSTEMS CORPORATION
- Product Code
- GEI
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- PHYSICIAN
Narratives
THE SUBJECT DEVICE WAS RETURNED TO OLYMPUS MEDICAL SYSTEMS CORPORATION (OMSC) FOR EVALUATION. THE EVALUATION CONFIRMED THAT THE PROBE BROKE DOWN AT 9.5 MM FROM THE DISTAL END. AND THERE WAS A SCRATCH AT THE BROKEN POINT. THERE WAS A CONTACT MARK OF THE PROBE AND GRASPING SECTION WHERE THE PROBE WAS BROKEN OFF. THE SHAPE OF THE FRACTURE SURFACE SHOWED THAT THE FRACTURE DEVELOPED FROM THE SCRATCHED AS THE STARTING POINT. THE MANUFACTURING HISTORY WAS REVIEWED, WITH NO IRREGULARITIES NOTED. BASED ON THE ANALYSIS RESULT ABOVE, SINCE THE DEVICE WAS GRASPING A THICK AND HARD TISSUE AND ACTIVATE WHILE TWISTING THE TISSUE, THE PROBE AND GRASPING SECTION CAME INTO CONTACT WITH EACH OTHER SPARKS OCCURRED BETWEEN THEM. THAT CAUSED THE DEVICE A SCRATCH OCCURRED. AFTER THAT, SINCE THE PHYSICIAN CONTINUED TO USE THE DEVICE, THE PROBE TIP BROKE. THE TB-0545PC INSTRUCTION MANUAL ALREADY STATES; WARNING: DO NOT APPLY ONLY THE PROBE TIP TO THE TISSUE WITH A STRONG FORCE OR PULL THE PROBE TIP UP GRASPING A TISSUE OR ACTIVATE OUTPUT WHILE TWISTING THE SHAFT OR TURNING THE ROTATION KNOB WITH THE TRANSDUCER AND CONNECTION CABLE. OTHERWISE, THE PROBE MAY BE DAMAGED BY INTERFERENCE WITH THE INTERNAL PARTS, WHICH COULD RESULT IN THE TIP BREAKING AND DROPPING INSIDE THE BODY CAVITY. THIS REPORT IS BEING SUBMITTED AS A MEDICAL DEVICE REPORT IN AN ABUNDANCE OF CAUTION.
WHEN A PHYSICIAN USED THE SUBJECT DEVICE FOR A RESECTION OF SIGMOID, THE PHYSICIAN DETECTED THE BROKEN PROBE BY LAPAROSCOPIC IMAGE. THE BROKEN PIECE OF THE PROBE WAS TAKEN OUT. THE PHYSICIAN REPLACED THE SUBJECT DEVICE WITH A SIMILAR DEVICE. THERE WAS NO PATIENT HARM REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 131263 | OLYMPUS THUNDERBEAT 5MM, 45CM, PISTOL GRIP | THUNDERBEAT HANDPIECE | GEI | OLYMPUS MEDICAL SYSTEMS CORPORATION | TB-0545PC | 22K |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |