FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 3060738 · Received April 5, 2013

Report

Report Number
8020893-2013-00789
Event Type
Malfunction
Date Received
April 5, 2013
Date of Event
March 13, 2013
Report Date
March 13, 2013
Manufacturer
COVIDIEN
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT AN 840 VENTILATOR WAS INOPERABLE. NO PATIENT INVOLVEMENT. THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION. THE CSE INSPECTED THE DEVICE AND REPLACED THE GUI CABLE. THE UNIT PASSED EXTENDED SELF-TESTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
139672 840 VENTILATOR VENTILATOR CBK COVIDIEN 840

Patients

Seq Age Sex Outcome Treatment
1