FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 3000 SYSTEM

MDR report key: 3060729 · Received March 29, 2013

Report

Report Number
3008642652-2013-00811
Event Type
Malfunction
Date Received
March 29, 2013
Date of Event
March 12, 2013
Report Date
March 20, 2013
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (DAMAGED CONNECTOR) WAS CONFIRMED. AS RECEIVED, THE TRUNK CABLE CONNECTOR HOUSING WAS BROKEN. THE CONNECTOR LOCKING NUT WAS MISSING, WHICH WOULD PREVENT THE ELECTRODE BELT FROM PROPERLY SECURING INTO THE MONITOR. THE ROOT CAUSE FOR THE MISSING LOCKING NUT CANNOT BE POSITIVELY IDENTIFIED BUT IS LIKELY PHYSICAL ABUSE. NO ADVERSE EVENT RESULTED FROM THE DAMAGED CABLE AND WIRES. THE PATIENT RECEIVED A REPLACEMENT ELECTRODE BELT.

Description of Event or Problem · 1

A (B)(6) MALE PATIENT'S DAUGHTER CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THAT THE ELECTRODE BELT TRUNK CABLE CONNECTOR WAS DAMAGED. THE PATIENT WAS ISSUED A REPLACEMENT ELECTRODE BELT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
129631 LIFEVEST WCD 3000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 3000 NA

Patients

Seq Age Sex Outcome Treatment
1 75 YR