FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 3060710 · Received April 5, 2013

Report

Report Number
8020893-2013-00817
Event Type
Malfunction
Date Received
April 5, 2013
Date of Event
January 20, 2013
Report Date
March 20, 2013
Manufacturer
COVIDIEN, FORMERLY NELLCO
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
NOT APPLICABLE

Narratives

Description of Event or Problem · 1

THE SERVICE REPORT SHOWS THE CUSTOMER REPORTED THAT THE 840 VENTILATOR STOPPED CYCLING WHILE IN USE ON A PT. THE PT WAS NOT HARMED OR INJURED AS A RESULT OF THE EVENT. THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION. THE CSE INSPECTED THE DEVICE AND REPLACED THE BREATH DELIVERY UNIT (BDU) CPU PCB. THE UNIT PASSED EXTENDED SELF-TESTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
139739 840 VENTILATOR VENTILATOR CBK COVIDIEN, FORMERLY NELLCO 840

Patients

Seq Age Sex Outcome Treatment
1