FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 3060699
·
Received March 29, 2013
Report
- Report Number
- 3008642652-2013-00822
- Event Type
- Malfunction
- Date Received
- March 29, 2013
- Date of Event
- January 1, 2013
- Report Date
- March 25, 2013
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF BATTERY PACK SN (B)(4) IS STILL UNDERWAY. THE REPORTED PROBLEM (FAULT ON CHARGER) WAS CONFIRMED. AS RECEIVED, THE BATTERY WOULD NOT CHARGE OR POWER ON A MONITOR. ROOT CAUSE ANALYSIS IS CURRENTLY UNDERWAY. A SUPPLEMENTAL REPORT WILL BE SENT UPON COMPLETION OF EVALUATION. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE BATTERY PACKS. THE PATIENT RECEIVED A REPLACEMENT BATTERY PACK.
Description of Event or Problem · 1
A (B)(6) MALE PATIENT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THAT ONE OF THE BATTERY PACK WAS FAULTING IN THE CHARGER. THE PATIENT WAS ISSUED A REPLACEMENT BATTERY PACK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 130855 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 64 YR |