FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 3060695 · Received March 29, 2013

Report

Report Number
3008642652-2013-00798
Event Type
Malfunction
Date Received
March 29, 2013
Date of Event
February 25, 2013
Report Date
March 20, 2013
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVAL SUMMARY: DEVICE EVAL OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (DAMAGED CABLE) HAS BEEN CONFIRMED. UPON EVAL, THE CABLE CONNECTING THE REAR THERAPY ELECTRODE (TE) AND THE DISTRIBUTION NODE (DN) WAS PULLED FROM THE DN STRAIN RELIEF, DAMAGING INTERNAL WIRES. THE ROOT CAUSE OF THE DAMAGED CABLE AND WIRES CANNOT BE POSITIVELY IDENTIFIED BUT IS LIKELY EXCESSIVE FORCE PLACED ON THE CABLE. NO ADVERSE EVENT RESULTED FROM THE DAMAGED CABLE AND WIRES. THE PT RECEIVED A REPLACEMENT ELECTRODE BELT.

Description of Event or Problem · 1

A (B)(6) MALE PT CONTACTED ZOLL CUSTOMER SUPPORT TO REPORT THAT ONE OF THE CABLES ON THE BELT WAS DAMAGED. THE PT WAS ISSUED A REPLACEMENT ELECTRODE BELT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
129641 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 70 YR