FDA Adverse Event
Malfunction
Summary report: N
HEARTSTART MRX - EMS DEFIBRILLATOR
MDR report key: 3060676
·
Received March 26, 2013
Report
- Report Number
- 1218950-2013-01048
- Event Type
- Malfunction
- Date Received
- March 26, 2013
- Report Date
- March 4, 2013
- Manufacturer
- PHILIPS MEDICAL SYSTEMS
- Product Code
- MKJ
- PMA / PMN Number
- K031187
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NV, US
- Reporter Occupation
- UNKNOWN
Narratives
Additional Manufacturer Narrative · 1
(B)(4): THE CUSTOMER REPORTED THAT DURING A 12 LEAD ECG THE V5 LEAD WAS INTERMITTENT. THERE IS NO REPORTED PT INVOLVEMENT. THE DEVICE WAS EVALUATED BY A PHILIPS FSE. THE REPORTED SYMPTOM WAS CONFIRMED. THE ISSUE WAS ISOLATED TO THE ECG TRUNK CABLE. THE ECG TRUNK CABLE WAS REPLACED TO RESOLVE THE ISSUE. THE DEVICE THEN PASSED ALL REQUIRED TESTING AND REMAINS AT THE CUSTOMER SITE FOR USE. THIS WAS A MALFUNCTION OF THE ECG TRUNK CABLE. REPLACEMENT OF THE TRUNK CABLE RESOLVED THE ISSUE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT DURING A 12 LEAD ECG THE V5 LEAD WAS INTERMITTENT. THERE IS NO REPORTED PT INVOLVEMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 124619 | HEARTSTART MRX - EMS DEFIBRILLATOR | MKJ | PHILIPS MEDICAL SYSTEMS | M3536A |