FDA Adverse Event Malfunction Summary report: N

HEARTSTART MRX - EMS DEFIBRILLATOR

MDR report key: 3060676 · Received March 26, 2013

Report

Report Number
1218950-2013-01048
Event Type
Malfunction
Date Received
March 26, 2013
Report Date
March 4, 2013
Manufacturer
PHILIPS MEDICAL SYSTEMS
Product Code
MKJ
PMA / PMN Number
K031187
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NV, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

(B)(4): THE CUSTOMER REPORTED THAT DURING A 12 LEAD ECG THE V5 LEAD WAS INTERMITTENT. THERE IS NO REPORTED PT INVOLVEMENT. THE DEVICE WAS EVALUATED BY A PHILIPS FSE. THE REPORTED SYMPTOM WAS CONFIRMED. THE ISSUE WAS ISOLATED TO THE ECG TRUNK CABLE. THE ECG TRUNK CABLE WAS REPLACED TO RESOLVE THE ISSUE. THE DEVICE THEN PASSED ALL REQUIRED TESTING AND REMAINS AT THE CUSTOMER SITE FOR USE. THIS WAS A MALFUNCTION OF THE ECG TRUNK CABLE. REPLACEMENT OF THE TRUNK CABLE RESOLVED THE ISSUE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT DURING A 12 LEAD ECG THE V5 LEAD WAS INTERMITTENT. THERE IS NO REPORTED PT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
124619 HEARTSTART MRX - EMS DEFIBRILLATOR MKJ PHILIPS MEDICAL SYSTEMS M3536A

Patients

Seq Age Sex Outcome Treatment
1