FDA Adverse Event Malfunction Summary report: N

SPECTRUM INFUSION PUMP

MDR report key: 3060585 · Received March 27, 2013

Report

Report Number
1314492-2013-00302
Event Type
Malfunction
Date Received
March 27, 2013
Date of Event
February 1, 2013
Report Date
February 27, 2013
Manufacturer
BAXTER HEALTHCARE CORPORATION
Product Code
FRN
PMA / PMN Number
K042121
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

MFR REF NO: (B)(4). THE DEVICE WAS NOT RETURNED TO BAXTER FOR EVAL AND THEREFORE AN EVAL COULD NOT BE COMPLETED. IF THE DEVICE IS RETURNED, AN EVAL WILL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE SUBMITTED. SYSTEM ERROR 322 WILL OCCUR WHEN THE PUMP TRANSITIONS FROM THE DOOR OPEN STATE TO THE DOOR CLOSED/SET-LOADED STATE AND THE LOWER LINK SWITCH DOES NOT ACTIVATE.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PUMP ALARMED FOR A SYSTEM ERROR 322 DURING TESTING. IT WAS ALSO REPORTED THAT THERE WAS NO PT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
124963 SPECTRUM INFUSION PUMP INFUSION PUMP FRN BAXTER HEALTHCARE CORPORATION 35700

Patients

Seq Age Sex Outcome Treatment
1