ANIMAS INSULIN INFUSION PUMP
Report
- Report Number
- 2531779-2013-04296
- Event Type
- Malfunction
- Date Received
- April 16, 2013
- Report Date
- March 20, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME. (B)(6).
FOLLOW-UP #1 DATE OF SUBMISSION (B)(4) 2013 DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: EVALUATION REVEALED THAT THE KEYPAD WAS INTACT. EVALUATION REVEALED THAT ALL BUTTONS RESPONDED APPROPRIATELY. THERE WAS NO EVIDENCE OF KEYPAD CONTAMINATION FOUND UNDER KEY CONTACTS. THE AUDIO BOLUS BUTTON COVER WAS INTACT. THE AUDIO BOLUS BUTTON WAS FOUND TO BE UNRESPONSIVE. THE AUDIO BOLUS BUTTON COVER WAS REMOVED AND THE WHITE SLUG TO BE MISSING. UNRELATED TO THE EVENT, THE DISPLAY LENS WAS SCRATCHED.
ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS, ALLEGING A BUTTON/KEYPAD (TACTILE CHANGES/UNRESPONSIVE) ISSUE. THE REPORTER STATED THAT THE AUDIO BOLUS BUTTON WAS UNRESPONSIVE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.