FDA Adverse Event Malfunction Summary report: N

9900

MDR report key: 3060469 · Received April 10, 2013

Report

Report Number
1720753-2013-04819
Event Type
Malfunction
Date Received
April 10, 2013
Date of Event
April 2, 2013
Report Date
April 10, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE INTERCONNECT CABLE AND ISD CIRCUIT BOARD WERE REPLACED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE C-ARM HAD NO POWER AND WOULD NOT BOOT UP. THERE IS NO REPORT OF PT INJURY ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
150161 9900 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9900

Patients

Seq Age Sex Outcome Treatment
1