FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 3060441 · Received April 10, 2013

Report

Report Number
1720753-2013-04829
Event Type
Malfunction
Date Received
April 10, 2013
Date of Event
March 26, 2013
Report Date
April 10, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE HIGH VOLTAGE TANK WAS REPLACED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE FIELD ENGINEER REPORTED THAT THERE WAS DISCHARGES IN THE HIGH VOLTAGE TANK AND THE SYSTEM WOULD NOT PERFORM FLUOROSCOPIC X-RAY. THERE IS NO REPORT OF PT INJURY ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
148488 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1