FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 3060438 · Received April 10, 2013

Report

Report Number
1720753-2013-04831
Event Type
Malfunction
Date Received
April 10, 2013
Date of Event
March 27, 2013
Report Date
April 10, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE PS2 POWER SUPPLY WAS REPLACED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE IMAGES WERE WHITE DOTS AND BLURRED, THEN THE SCREEN WAS BLACK WITH NO IMAGE. THIS RESULTED IN A LOSS OF THE LIVE IMAGE. THERE IS NO REPORT OF PT INJURY ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
148392 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1