FDA Adverse Event Malfunction Summary report: N

8800

MDR report key: 3060389 · Received April 10, 2013

Report

Report Number
1720753-2013-04860
Event Type
Malfunction
Date Received
April 10, 2013
Date of Event
March 27, 2013
Report Date
April 10, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE X-RAY MONITOR WAS REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

IT WAS REPORTED THE SYSTEM DISPLAYED A GENERATOR ERROR. THIS ERROR IS LIKELY TO RESULT IN AN UN-COMMANDED LOSS OF SYSTEM FUNCTIONALITY AND REQUIRE A SYSTEM REBOOT. THERE IS NO REPORT OF INJURY OR DEATH.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
148144 8800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 8800

Patients

Seq Age Sex Outcome Treatment
1