FDA Adverse Event
Malfunction
Summary report: N
6800
MDR report key: 3060387
·
Received April 10, 2013
Report
- Report Number
- 1720753-2013-04861
- Event Type
- Malfunction
- Date Received
- April 10, 2013
- Date of Event
- March 21, 2013
- Report Date
- April 10, 2013
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE CPU WAS REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE SYSTEM WOULD INTERMITTENTLY NOT BOOT UP. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THE EVENT DESCRIBED IN THIS COMPLAINT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 149994 | 6800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 6800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |