FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 3060358 · Received April 11, 2013

Report

Report Number
1720753-2013-04874
Event Type
Malfunction
Date Received
April 11, 2013
Date of Event
March 12, 2013
Report Date
April 11, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TW
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE INTERCONNECT CABLE WAS REPLACED. THE SYSTEM WAS THEN FOUND TO BE OPERATING AS INTENDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SCREEN WAS INTERMITTENTLY GOING BLACK. THIS WOULD RESULT IN A LOSS OF THE LIVE IMAGE. THERE IS NO REPORT OF PATIENT INJURY ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
156848 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1