FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 3060353 · Received April 11, 2013

Report

Report Number
1720753-2013-04879
Event Type
Malfunction
Date Received
April 11, 2013
Date of Event
April 4, 2013
Report Date
April 11, 2013
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ON SITE INVESTIGATION. THE DISPLAY ADAPTOR BOARD WAS REPLACED DURING THE SERVICE CALL. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE LEFT MONITOR WAS BLANK. THIS WILL CAUSE THE SYSTEM TO BE UNUSABLE DUE TO A LOSS OF THE LIVE IMAGE. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
154613 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1