FDA Adverse Event
Injury
Summary report: N
NICOLET ONE MONITOR ON ERGO CART
MDR report key: 3060308
·
Received April 11, 2013
Report
- Report Number
- 3008289288-2013-00003
- Event Type
- Injury
- Date Received
- April 11, 2013
- Date of Event
- March 22, 2013
- Report Date
- April 9, 2013
- Manufacturer
- NATUS NEUROLOGY INCORPORATED
- Product Code
- OLV
- PMA / PMN Number
- K021185
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Additional Manufacturer Narrative · 1
WHEN PRODUCT IS SHIPPED TO THE CUSTOMER, THERE ARE SEVERAL SETS OF SCREWS AND INSTRUCTIONS EXPLAINING WHICH SCREWS TO USE WHERE. THE CUSTOMER INSERTED INCORRECT SHORTER SCREWS WHICH CAUSED THE PC TO FALL. WE HAVE ADDED CLEARER INSTALLATION INSTRUCTIONS STARTING IN (B)(6) 2013 TO SYSTEMS SHIPPED.
Description of Event or Problem · 1
PANEL PERSONAL COMPUTER FELL OFF MOUNT, HANGING FROM CABLES. THERE WAS NO PATIENT OR USER INJURY, BUT IF THEY WERE BELOW THE COMPUTER WHEN IT FELL, THEY COULD HAVE BEEN HURT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 154304 | NICOLET ONE MONITOR ON ERGO CART | NICOLET ONE MONITOR ON ERGO CART | OLV | NATUS NEUROLOGY INCORPORATED | 982A0500 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Other |