FDA Adverse Event Other Summary report: N

W&H WA-99LT 1:5 HIGH-SPEED ATTACHMENT

MDR report key: 3059679 · Received April 11, 2013

Report

Report Number
3015729-2013-00001
Event Type
Other
Date Received
April 11, 2013
Date of Event
April 8, 2013
Report Date
April 9, 2013
Manufacturer
W&H DENTALWERK BURMOOS GMBH
Product Code
EFB
PMA / PMN Number
K070663
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
DENTIST

Narratives

Additional Manufacturer Narrative · 1

#(B)(4) WAS LAST REPAIRED (B)(4) 2008, #(B)(4) WAS LAST REPAIRED (B)(4) 2008. BOTH ATTACHMENTS ARE (B)(4). THERE ARE NOTES IN THE PREVIOUS REPAIRS THAT THERE WERE POSSIBLY GENERIC REPAIR PARTS IN THE ATTACHMENTS. GENERATION 1 HANDPIECE ATTACHMENTS ARE NO LONGER REPAIRED BUT ARE REPLACED WITH GENERATION 3 DEVICES. (B)(4).

Description of Event or Problem · 1

WHILE PERFORMING A DENTAL PROCEDURE, HANDPIECE ATTACHMENT OVERHEATED AND BURNED PATIENT. PATIENT WAS REFERRED TO A PLASTIC SURGEON FOR TREATMENT OF BURN ON TONGUE. THIS WAS THE FIRST PATIENT BURN REPORTED BY THE OFFICE, ORIGINALLY REPORTED (B)(4) 2012, BUT FOLLOW-UP TREATMENT WAS NOT INDICATED AT THAT TIME. SUBSEQUENTLY, A SECOND HANDPIECE ATTACHMENT, S/N (B)(4), BECAME THE CAUSE FOR THIS CURRENT EVENT. THE CURRENT EVENT WAS A BURNED LIP THAT DID NOT REQUIRE INTERVENTION OR FOLLOW-UP TREATMENT AND IS NOT DIRECTLY THE SUBJECT OF THIS REPORT; RATHER THE METHOD BY WHICH WE BECAME AWARE THE FIRST INCIDENT WAS REPORTABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
155214 W&H WA-99LT 1:5 HIGH-SPEED ATTACHMENT SPEED-INCREASING HANDPIECE ATTACHMENT EFB W&H DENTALWERK BURMOOS GMBH WA-99LT

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention