W&H WA-99LT 1:5 HIGH-SPEED ATTACHMENT
Report
- Report Number
- 3015729-2013-00001
- Event Type
- Other
- Date Received
- April 11, 2013
- Date of Event
- April 8, 2013
- Report Date
- April 9, 2013
- Manufacturer
- W&H DENTALWERK BURMOOS GMBH
- Product Code
- EFB
- PMA / PMN Number
- K070663
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- DENTIST
Narratives
#(B)(4) WAS LAST REPAIRED (B)(4) 2008, #(B)(4) WAS LAST REPAIRED (B)(4) 2008. BOTH ATTACHMENTS ARE (B)(4). THERE ARE NOTES IN THE PREVIOUS REPAIRS THAT THERE WERE POSSIBLY GENERIC REPAIR PARTS IN THE ATTACHMENTS. GENERATION 1 HANDPIECE ATTACHMENTS ARE NO LONGER REPAIRED BUT ARE REPLACED WITH GENERATION 3 DEVICES. (B)(4).
WHILE PERFORMING A DENTAL PROCEDURE, HANDPIECE ATTACHMENT OVERHEATED AND BURNED PATIENT. PATIENT WAS REFERRED TO A PLASTIC SURGEON FOR TREATMENT OF BURN ON TONGUE. THIS WAS THE FIRST PATIENT BURN REPORTED BY THE OFFICE, ORIGINALLY REPORTED (B)(4) 2012, BUT FOLLOW-UP TREATMENT WAS NOT INDICATED AT THAT TIME. SUBSEQUENTLY, A SECOND HANDPIECE ATTACHMENT, S/N (B)(4), BECAME THE CAUSE FOR THIS CURRENT EVENT. THE CURRENT EVENT WAS A BURNED LIP THAT DID NOT REQUIRE INTERVENTION OR FOLLOW-UP TREATMENT AND IS NOT DIRECTLY THE SUBJECT OF THIS REPORT; RATHER THE METHOD BY WHICH WE BECAME AWARE THE FIRST INCIDENT WAS REPORTABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 155214 | W&H WA-99LT 1:5 HIGH-SPEED ATTACHMENT | SPEED-INCREASING HANDPIECE ATTACHMENT | EFB | W&H DENTALWERK BURMOOS GMBH | WA-99LT |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |