PRECISION FLOW
Report
- Report Number
- 1125759-2013-00002
- Event Type
- Malfunction
- Date Received
- March 22, 2013
- Date of Event
- January 30, 2013
- Report Date
- March 21, 2013
- Manufacturer
- VAPOTHERM, INC
- Product Code
- BTT
- PMA / PMN Number
- K072845
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
A PRECISION FLOW UNIT IN (B)(6) WAS REPORTED TO HAVE BEEN EXPERIENCING FLOW RATES THAT WERE HIGHER THAN THE DIALED IN SETTING. THE HIGH FLOWS COULD CONTINUE FOR UP TO 15 SECONDS BEFORE AND GENERAL FAULT ALARM WOULD BE TRIGGERED. THERE WERE NO PT INJURIES AS A RESULT HOWEVER 15 SECONDS OF HIGHER THAN DESIRED FLOW RATES COULD POTENTIALLY CONTRIBUTE TO A PT INJURY IF IT WERE TO OCCUR AGAIN. THE BLOCKED TUBE ALARM IS DESIGNED TO PREVENT THE FLOW RATE FROM INCREASING MORE THAN 10% ABOVE THE DIAL IN FLOW RATE. HOWEVER, IN THIS CASE THE BLOCKED TUBE ALARM FUNCTION HAD BEEN TURNED OFF DURING A PREVIOUS SERVICING OF THE DEVICE AT A SERVICE CENTER SELECTED BY THE DISTRIBUTOR. THIS FUNCTION SHOULD HAVE BEEN TURNED BACK ON PRIOR TO SENDING THE UNIT BACK TO THE CUSTOMER, BUT THE PROCESS STEP WAS MISSED BY THE SERVICE TECHNICIAN. IT SHOULD BE NOTED THAT IT IS NOT POSSIBLE TO TURN ALL ALARM FUNCTIONS OFF. THE SERVICE CENTER HAS MODIFIED THEIR PROCEDURES AND A CAPA HAS BEEN INITIATED BY THE MFR TO PREVENT THIS ERROR FROM HAPPENING AGAIN.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 120931 | PRECISION FLOW | PRECISION FLOW | BTT | VAPOTHERM, INC | PF-UNIT |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |