COMPACT AIR DRIVE II
Report
- Report Number
- 8030965-2013-01652
- Event Type
- Malfunction
- Date Received
- April 15, 2013
- Date of Event
- March 18, 2013
- Report Date
- March 19, 2013
- Manufacturer
- SYNTHES GMBH
- Product Code
- HWE
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. A SERVICE HISTORY OF THE PAST SIX MONTHS HAS BEEN REVIEWED. NO SERVICE HISTORY REVIEW CAN BE PERFORMED. THE ITEM HAS NOT PREVIOUSLY BEEN IN FOR SERVICE. THERE IS NO INFORMATION RELEVANT TO THE CURRENT COMPLAINED ISSUE. (B)(4). PLACEHOLDER.
DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. DEVICE WAS RECEIVED FOR EVALUATION. INVESTIGATION COORDINATED BY SYNTHES (B)(4). REPORT RECEIVED INDICATES THE REPORTERS COMPLAINT THAT THE COMPACT AIR DRIVE DRILL IS OVERHEATING COULD NOT BE CONFIRMED. PLACEHOLDER.
DEVICE USED FOR TREATMENT AND NOT DIAGNOSIS. THE MANUFACTURING DOCUMENTS WERE REVIEWED AND NO COMPLAINT RELATED ISSUES WERE FOUND.
DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. DEVICE IS AN INSTRUMENT AND IS NOT IMPLANTED/EXPLANTED. THE INVESTIGATION COULD NOT BE COMPLETED; NO CONCLUSION COULD BE DRAWN, AS NO PRODUCT WAS RECEIVED. A REVIEW OF THE DEVICE HISTORY RECORDS HAS BEEN REQUESTED.
DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. SYNTHES LOT NUMBER IS 4951162.
THE USER FACILITY REPORTS DURING TESTING ON (B)(6) 2013, THE COMPACT AIR DRIVE WAS OVERHEATING. DEVICE WAS NOT USED IN SURGICAL PROCEDURE, NO PATIENT INVOLVEMENT WAS REPORTED. THIS IS 1 OF 1 REPORT FOR COMPLAINT (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 160936 | COMPACT AIR DRIVE II | HWE | SYNTHES GMBH | 4951162 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |