FDA Adverse Event Other Summary report: N

*

MDR report key: 305639 · Received November 17, 2000

Report

Report Number
305639
Event Type
Other
Date Received
November 17, 2000
Date of Event
October 4, 2000
Report Date
November 5, 2000
Manufacturer
INFIMED, INC.
Product Code
JAA
Adverse Event
Yes
Report Source
User Facility report
Reporter Occupation
INVALID DATA

Narratives

Description of Event or Problem · 1

VIDEO MONITOR FOR FLOUROSCOPY MOUNTED ON PIECE OF EQUIP. (ARM FROM CEILING) - TO BE ALLOWED TO SWIVEL. NO SUPPORT FOR MONITOR FROM BOTTOM. WEIGHT OF MONITOR HANGS FROM BRACKET MOUNTED ONTO TOP ONLY. TECH WAS SWIVELING MONITOR TO SEE SCREEN. PT LYING ON TABLE, BELOW. MONITOR BROKE FROM BRACKET, WAS FALLING, TECH TRIED TO CATCH IT TO KEEP IT FROM INJURING PT, HIT THE TECH IN THE FOREHEAD, HIT THE PT ON THE RIGHT ARM ABOVE ELBOW - ONLY INJURY - SMALL BRUISE - SORENESS TO PT. TECH - SMALL LACERATION, BRUISE, HEADACHE.

Description of Event or Problem · 2

VIDOE MONITOR FOR FLUROSCOPY MOUNTED ON PIECE OF EQUIP. (ARM FROM CEILING)- TO BE ALLOWED TO SWIVEL. NO SUPPORT FOR MONITOR FROM BOTTOM. WEIGHT OF MONITOR HANGS FROM BRACKET MOUNTED ONTO TOP ONLY. TECH WAS SWIVELING MONITOR TO SEE SCREEN. PT LYING ON TABLE BELOW. MONITOR BROKE FROM BRACKET, WAS FALLING, TECH TRIED TO CATCH IT TO KEEP IT FROM INJURING PT, HIT THE TECH IN THE FOREHEAD, HIT THE PT ON THE RIGHT ARM ABOVE ELBOW-ONLY INJURY-SMALL BRUISE-SORENESS TO PT. TECH-SMALL LACERATION, BRUISE, HEADACHE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 * * JAA INFIMED, INC. * *

Patients

Seq Age Sex Outcome Treatment
1 50 YR Other
2 UNKNOWN Other