FDA Adverse Event Injury Summary report: N

DEXTRUS 4136

MDR report key: 3053642 · Received April 11, 2013

Report

Report Number
1028232-2013-00929
Event Type
Injury
Date Received
April 11, 2013
Date of Event
February 20, 2013
Report Date
March 28, 2013
Manufacturer
BIOTRONIK SE & CO. KG
Product Code
NVN
PMA / PMN Number
P950037
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT, THE LEAD WAS SUBJECTED TO AN EXTENSIVE ANALYSIS. THE PERFORMANCE OF THE DEVICE UNDER COMPLAINT WAS SCRUTINIZED, INCLUDING A VISUAL, ELECTRICAL AND MECHANICAL INSPECTION. THE VISUAL INSPECTION SHOWED CUTTINGS IN THE INSULATION AND DEFORMATIONS OF THE CONDUCTOR COIL WHICH OCCURRED MOST LIKELY DURING SURGERY. IN THE COURSE OF THE FURTHER ANALYSIS, NO DEVIATIONS WERE NOTED WHICH MIGHT BE RELATED TO THE CLINICAL OBSERVATION AS MENTIONED IN THE COMPLAINT DESCRIPTION. THE LEAD PROVED TO BE WITHIN SPECIFICATIONS AND FLAWLESS THROUGHOUT ITS INSPECTION. IN SUMMARY, THERE WAS NO SIGN OF A MATERIAL OR MANUFACTURING PROBLEM.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT ONE DAY POST IMPLANT THIS RIGHT ATRIAL LEAD EXHIBITED UNDERSENSING. IT WAS DETERMINED THE LEAD WAS DISLODGED. SURGICAL INTERVENTION WAS PERFORMED AND THIS LEAD WAS EXPLANTED AND REPLACED. IT WAS REPORTED THE PATIENT WAS DOING FINE WITH NO ADVERSE PATIENT EFFECTS NOTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
154467 DEXTRUS 4136 PACER LEAD NVN BIOTRONIK SE & CO. KG 358754

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization