Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT, THE LEAD WAS SUBJECTED TO AN EXTENSIVE ANALYSIS. THE PERFORMANCE OF THE DEVICE UNDER COMPLAINT WAS SCRUTINIZED, INCLUDING A VISUAL, ELECTRICAL AND MECHANICAL INSPECTION. THE VISUAL INSPECTION SHOWED CUTTINGS IN THE INSULATION AND DEFORMATIONS OF THE CONDUCTOR COIL WHICH OCCURRED MOST LIKELY DURING SURGERY. DURING FURTHER ANALYSIS, NO DEVIATIONS WERE NOTED WHICH MIGHT BE RELATED TO THE CLINICAL OBSERVATION AS MENTIONED IN THE COMPLAINT DESCRIPTION. THE LEAD PROVED TO BE WITHIN SPECIFICATIONS AND FLAWLESS THROUGHOUT ITS INSPECTION. IN SUMMARY, THERE WAS NO SIGN OF A MATERIAL OR MANUFACTURING PROBLEM.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS RIGHT ATRIAL LEAD DISLODGED. A REVISION PROCEDURE WAS PERFORMED AND THE LEAD WAS EXPLANTED. NO ADVERSE PATIENT EFFECTS WERE REPORTED DURING THE PROCEDURE. THE LEAD WAS RETURNED FOR RELIABILITY ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
156628 DEXTRUS 4135 PACER LEAD NVN BIOTRONIK SE & CO. KG 358744

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization