FLEXTEND
Report
- Report Number
- 2124215-2013-05250
- Event Type
- Injury
- Date Received
- April 11, 2013
- Date of Event
- March 27, 2013
- Report Date
- April 19, 2013
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- NVN
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AZ
- Reporter Occupation
- PHYSICIAN
Narratives
AS OF TODAY, THE LEAD HAS NOT BEEN RETURNED FOR ANALYSIS. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS REPORT WILL BE UPDATED.
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE LEAD WAS PERFORMED. RESISTANCE AND PRESSURE TESTS WERE COMPLETED TO ASSESS LEAD ELECTRICAL PERFORMANCE AND INSULATION INTEGRITY. MEASUREMENTS THROUGHOUT THESE TESTS WERE WITHIN NORMAL LIMITS. MICROSCOPIC INSPECTIONS OF THE TERMINAL PIN ASSEMBLY, LEAD BODY, AND ELECTRODE TIP FOUND NO ANOMALIES. LABORATORY TESTING WAS UNABLE TO REPRODUCE THE REPORTED CLINICAL OBSERVATIONS AND DETAILED ANALYSIS DID NOT REVEAL ANY ABNORMALITIES.
THE LEAD HAS BEEN RETURNED FOR ANALYSIS.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS RIGHT ATRIAL (RA) LEAD WAS NOT CAPTURING. IT WAS DETERMINED THAT THE LEAD HAD DISLODGED. THE PATIENT UNDERWENT A LEAD REVISION PROCEDURE WHERE THE PHYSICIAN NOTED THAT THE LEAD WAS WRAPPED AROUND THE PULSE GENERATOR. THE PHYSICIAN BELIEVED THAT THE PATIENT WAS A TWIDDLER. THE LEAD WAS EXPLANTED AND REPLACED. THERE WERE NO ADVERSE PATIENT EFFECTS REPORTED. THE LEAD IS TO BE RETURNED FOR ANALYSIS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 155870 | FLEXTEND | IMPLANTABLE LEAD | NVN | CPI - DEL CARIBE | 4087 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 73 YR | Hospitalization| L| R | E142| 4087| 0295 |