FDA Adverse Event Injury Summary report: N

FLEXTEND

MDR report key: 3053612 · Received April 11, 2013

Report

Report Number
2124215-2013-05250
Event Type
Injury
Date Received
April 11, 2013
Date of Event
March 27, 2013
Report Date
April 19, 2013
Manufacturer
CPI - DEL CARIBE
Product Code
NVN
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
AZ
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AS OF TODAY, THE LEAD HAS NOT BEEN RETURNED FOR ANALYSIS. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS REPORT WILL BE UPDATED.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE LEAD WAS PERFORMED. RESISTANCE AND PRESSURE TESTS WERE COMPLETED TO ASSESS LEAD ELECTRICAL PERFORMANCE AND INSULATION INTEGRITY. MEASUREMENTS THROUGHOUT THESE TESTS WERE WITHIN NORMAL LIMITS. MICROSCOPIC INSPECTIONS OF THE TERMINAL PIN ASSEMBLY, LEAD BODY, AND ELECTRODE TIP FOUND NO ANOMALIES. LABORATORY TESTING WAS UNABLE TO REPRODUCE THE REPORTED CLINICAL OBSERVATIONS AND DETAILED ANALYSIS DID NOT REVEAL ANY ABNORMALITIES.

Description of Event or Problem · 1

THE LEAD HAS BEEN RETURNED FOR ANALYSIS.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS RIGHT ATRIAL (RA) LEAD WAS NOT CAPTURING. IT WAS DETERMINED THAT THE LEAD HAD DISLODGED. THE PATIENT UNDERWENT A LEAD REVISION PROCEDURE WHERE THE PHYSICIAN NOTED THAT THE LEAD WAS WRAPPED AROUND THE PULSE GENERATOR. THE PHYSICIAN BELIEVED THAT THE PATIENT WAS A TWIDDLER. THE LEAD WAS EXPLANTED AND REPLACED. THERE WERE NO ADVERSE PATIENT EFFECTS REPORTED. THE LEAD IS TO BE RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
155870 FLEXTEND IMPLANTABLE LEAD NVN CPI - DEL CARIBE 4087

Patients

Seq Age Sex Outcome Treatment
1 73 YR Hospitalization| L| R E142| 4087| 0295