FDA Adverse Event Injury Summary report: N

INSIGNIA

MDR report key: 3053608 · Received April 11, 2013

Report

Report Number
2124215-2013-05378
Event Type
Injury
Date Received
April 11, 2013
Date of Event
September 3, 2012
Report Date
January 6, 2014
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
NVZ
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCT IS EXPECTED TO BE RETURNED FOR ANALYSIS. THIS REPORT WILL BE UPDATED UPON RETURN AND COMPLETION OF ANALYSIS.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, DETAILED MECHANICAL AND ELECTRICAL TESTING WAS PERFORMED ON THE DEVICE. THE DEVICE BATTERY STATUS WAS EOL. THE DEVICE FUNCTIONED NORMALLY THROUGHOUT TESTING. LABORATORY ANALYSIS DETERMINED THAT THIS DEVICE EXPERIENCED NORMAL BATTERY DEPLETION; HOWEVER, THE ESTIMATED LONGEVITY REMAINING VALUE APPEARED TO DECREASE MORE QUICKLY THAN EXPECTED BETWEEN ROUTINE FOLLOW-UPS. FACTORS INFLUENCING THE ESTIMATED LONGEVITY REMAINING CALCULATION INCLUDE PACING RATE, AMPLITUDE, PULSE-WIDTH AND LEAD IMPEDANCE. ANY (EVEN SLIGHT) CHANGES IN THESE FACTORS WILL IMPACT THE BATTERY CONSUMPTION CALCULATION AND THEREFORE THE REMAINING LONGEVITY ESTIMATE. PLEASE NOTE THAT, DESPITE THE DROP IN ESTIMATED LONGEVITY REMAINING, THE ACTUAL BATTERY CONDITION DID NOT CHANGE SIGNIFICANTLY BETWEEN FOLLOW-UPS. IN SUMMARY, IT WAS DETERMINED THAT THIS DEVICE EXPERIENCED NORMAL BATTERY DEPLETION, BUT DECLARED EOL EARLIER THAN PREVIOUSLY ESTIMATED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS PACEMAKER DECLARED END OF LIFE (EOL) BATTERY STATUS EARLIER THAN EXPECTED. THE MAGNET RATE WAS 100 BPM AT THE DEVICE CHECK IN (B)(6) 2011. IN (B)(6) 2012, THE MAGNET RATE WAS 90 BPM AND THE REMAINING LONGEVITY WAS 1 YEAR. HOWEVER, WHEN THE DEVICE WAS CHECKED IN (B)(6) 2012, IT WAS NOTED THAT EOL HAD BEEN DECLARED IN (B)(6) 2012. THE DEVICE WAS THEN EXPLANTED AND REPLACED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
154278 INSIGNIA IMPLANTABLE PULSE GENERATOR NVZ GUIDANT CRM CLONMEL IRELAND 1292

Patients

Seq Age Sex Outcome Treatment
1 91 YR Hospitalization| L| R 1298| 435-08| 294-03| 431-07| 1292