FDA Adverse Event Injury Summary report: N

ANIMAS INSULIN CARTRIDGE

MDR report key: 3053519 · Received April 11, 2013

Report

Report Number
2531779-2013-04083
Event Type
Injury
Date Received
April 11, 2013
Date of Event
April 6, 2013
Report Date
April 6, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K032257
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE CARTRIDGE HAS NOT BEEN RETURNED TO ANIMAS. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Additional Manufacturer Narrative · 1

(B)(4) - DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: THE CARTRIDGE WAS RETURNED TO ANIMAS. A VISUAL INSPECTION OF THE CARTRIDGE WAS PERFORMED. NO DAMAGE OR DEFECTS WERE NOTED. A LEAK TEST WAS PERFORMED WITH NO FAILURES OBSERVED. EACH CARTRIDGE LOT IS SUBJECTED TO A STATISTICAL SAMPLING PLAN AND MUST PASS TESTING FOR FORCE, CRACKS, AND FOREIGN MATERIAL PRIOR TO RELEASE FOR DISTRIBUTION.

Description of Event or Problem · 1

THE PATIENT CONTACTED ANIMAS ON (B)(4) 2013 REPORTING BLOOD GLUCOSE LEVELS UP TO 32.2 MMOL/L WITH INCREASED THIRST, VOMITING, AND UPSET STOMACH. THE PATIENT REPORTED THAT AFTER MULTIPLE SITE CHANGES, BLOOD GLUCOSE LEVELS WERE DOWN TO 25.4 MMOL/L . THE PATIENT REPORTED THAT UPON WAKING IN THE MORNING, THE TUBING WAS FOUND TO BE DISCONNECTED FROM THE CARTRIDGE. THE PUMP PRIME HISTORY INDICATED THAT THE LAST SITE CHANGE WAS TWO NIGHTS PRIOR. THE PATIENT CONFIRMED THAT THE LUER CONNECTION WAS NOT BROKEN AND THE CARTRIDGE AND TUBING WERE WORKING FOR SOME TIME PRIOR TO BECOMING DISCONNECTED. THE PATIENT CONFIRMED THAT THE CONNECTION WAS TIGHT WHEN THE CARTRIDGE WAS PLACED INTO THE PUMP. THIS REPORT IS MADE BASED ON THE ALLEGATION THAT THE PATIENT EXPERIENCED HYPERGLYCEMIA RELATED TO THE LUER CONNECTION BECOMING DISCONNECTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
156111 ANIMAS INSULIN CARTRIDGE INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 67 YR Life Threatening