INTERMATE
Report
- Report Number
- 1416980-2013-09052
- Event Type
- Malfunction
- Date Received
- April 11, 2013
- Report Date
- March 20, 2013
- Manufacturer
- BAXTER HEALTHCARE - IRVINE
- Product Code
- FRN
- PMA / PMN Number
- K910425
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FR
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND IS CURRENTLY IN THE PROCESS OF BEING EVALUATED. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL RELEVANT DETAILS BECOME AVAILABLE. THE BATCH REVIEW REVEALED THAT ALL OF THE ACCEPTANCE CRITERIA WERE MET TO RELEASE THE LOT. THERE WERE NO NON-CONFORMANCES, FAILURES, REWORK, OR DEVIATIONS RELATED TO THE LOT. THERE WERE NO CHANGES TO SPECIFICATIONS, TEST METHODS, PROCESS, EQUIPMENT, OR RAW MATERIALS THAT COULD BE ASSOCIATED WITH THE REPORTED CONDITION.
(B)(4). EVALUATION SUMMARY: BAXTER RECEIVED ONE SAMPLE FOR EVALUATION. THE REPORTED CONDITION OF A RUPTURED RESERVOIR WAS CONFIRMED. A VISUAL EXAMINATION OF THE UNIT DETERMINED THAT THE BLADDER RUPTURED IN A FOOTED POSITION. THE ROOT CAUSE COULD NOT BE DETERMINED. NO REPAIR WAS DONE, AS THIS IS A SINGLE-USE DEVICE WHICH WILL BE DISCARDED. NO OTHER OBSERVATIONS WERE NOTED ON THE UNIT.
IT WAS REPORTED THAT AN INTERMATE HAD A RESERVOIR RUPTURE AT THE END OF INFUSION. THE DEVICE WAS FILLED WITH A SOLUTION OF TICARPEN. THERE WAS PATIENT INVOLVEMENT, BUT THERE WAS NO REPORT OF PATIENT INJURY, MEDICAL INTERVENTION NECESSARY, OR ADVERSE REACTION IN ASSOCIATION WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.