FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 3053394 · Received April 11, 2013

Report

Report Number
1416980-2013-09043
Event Type
Malfunction
Date Received
April 11, 2013
Date of Event
November 1, 2012
Report Date
April 2, 2013
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). DEVICE EVALUATION: THE REPORTED CONDITION OF "DID NOT FUNCTION" WAS NOT CONFIRMED DURING SERVICE EVALUATION; HOWEVER, THE QUALITY ENGINEER CONFIRMED THE REPORTED CONDITION AS FAILURE CODE 94. THE EVALUATION WAS PERFORMED ONSITE BY A FIELD SERVICE TECHNICIAN. THE BATTERY WAS FOUND TO BE DISCHARGED WHICH RESULTED IN THE DATE AND HOUR DATA BEING DELETED. THE DATE AND HOUR DATA WAS RECONFIGURED TO CORRECT THE REPORTED CONDITION. A SERVICE HISTORY WAS PERFORMED REVEALING NO PREVIOUS SERVICE FOR THE REPORTED CONDITION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP MDR WILL BE SUBMITTED.

Description of Event or Problem · 1

THE FACILITY REPRESENTATIVE REPORTED A FLOGARD INFUSION PUMP THAT DID NOT FUNCTION. UPON FURTHER INVESTIGATION BY THE QUALITY ENGINEER, THE REPORTED EVENT WAS CONFIRMED AS A FAILURE CODE 94. IT IS UNKNOWN AT WHICH PROCESS STEP OR IN WHICH CARE AREA THIS EVENT OCCURRED. IT IS UNKNOWN IF THERE WAS ANY PATIENT INVOLVEMENT, HOWEVER, THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION FOR THE REPORTED CONDITION. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
154222 FLOGARD PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1