FDA Adverse Event Malfunction Summary report: N

ENERGEN

MDR report key: 3053296 · Received April 11, 2013

Report

Report Number
2124215-2013-04853
Event Type
Malfunction
Date Received
April 11, 2013
Date of Event
March 13, 2013
Report Date
April 4, 2013
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE LOCAL AREA FIELD REPRESENTATIVE WAS CONTACTED FOR ADDITIONAL INFORMATION. AT THIS TIME, NO FURTHER INFORMATION IS AVAILABLE AND RECORDS INDICATE THESE PRODUCTS REMAIN IN SERVICE. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE THIS REPORT WILL BE UPDATED.

Additional Manufacturer Narrative · 1

AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS IMPLANTABLE DEFIBRILLATION LEAD AND IMPLANTABLE CARDIOVERTER DEFIBRILLATOR (ICD) EXHIBITED A HIGH OUT OF RANGE SHOCK IMPEDANCE MEASUREMENT. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Description of Event or Problem · 1

ADDITIONAL INFORMATION WAS RECEIVED INDICATING THE PATIENT WAS SEEN IN CLINIC. SHOCK IMPEDANCE MEASUREMENTS IN CLINIC WERE 106 OHMS. IT WAS REPORTED THE PATIENT WAS UNDERGOING RADIATION AT THE TIME THE HIGH OUT OF RANGE MEASUREMENT WAS RECORDED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
157010 ENERGEN IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND E143

Patients

Seq Age Sex Outcome Treatment
1 63 YR 4469| E143| 0180