FDA Adverse Event Malfunction Summary report: N

EASYTRAK 2

MDR report key: 3053254 · Received April 11, 2013

Report

Report Number
2124215-2013-05095
Event Type
Malfunction
Date Received
April 11, 2013
Date of Event
March 11, 2013
Report Date
April 10, 2013
Manufacturer
CPI - DEL CARIBE
Product Code
OJX
PMA / PMN Number
P010012
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, THIS LEAD WAS CONFIRMED TO BE FRACTURED 31.6 TO 33.6 MM FROM THE TERMINAL PIN. BASED ON THE ANALYSIS RESULTS, WE SUSPECT THAT FATIGUE OR STRESS IN THE REGION OF THE FRACTURE SITE DUE TO RELATIVE MOTION BETWEEN THE SUTURE SLEEVE AND AN ANATOMICAL FEATURE LED TO THE FRACTURE. AS A LEAD MOVES IN RESPONSE TO NORMAL HEART RHYTHMS AND BLOOD FLOW, EXTENSIVE FLEXING OVER A PERIOD OF TIME MAY CAUSE FATIGUE OR STRESS, WEAKENING THE COIL, ULTIMATELY RESULTING IN A FRACTURE. THIS CAN OCCUR BETWEEN THE SUTURE SLEEVE AND SOME OTHER PART OF THE ANATOMY. FATIGUE FRACTURES IN THE POCKET OR CLAVICULAR/FIRST RIB AREA ARE WELL KNOWN AND DOCUMENTED IN THE INDUSTRY. A COMBINATION OF LEAD DESIGN, IMPLANT TECHNIQUES, AND PATIENT ANATOMY AND ACTIVITY LEVEL CONTRIBUTE TO THESE TYPES OF OCCURRENCES.

Additional Manufacturer Narrative · 1

THE LEAD WAS RETURNED AND IS CURRENTLY IN ANALYSIS. UPON COMPLETION FROM ANALYSIS, THIS REPORT WILL BE UPDATED.

Description of Event or Problem · 1

--

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS LEFT VENTRICULAR (LV) LEAD WAS SUSPECTED TO BE FRACTURED VIA CHEST X-RAY. IMPEDANCE MEASUREMENTS WERE OUT OF RANGE AND GREATER THAN 2,000 OHMS IN ALL PACING VECTORS. THE LEAD WAS EXPLANTED DURING A PULSE GENERATOR REPLACEMENT PROCEDURE. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
155409 EASYTRAK 2 IMPLANTABLE LEAD OJX CPI - DEL CARIBE 4542

Patients

Seq Age Sex Outcome Treatment
1 61 YR 4542| 0144| 1188T| T125| 1831| 4548| N119