FDA Adverse Event Malfunction Summary report: N

TELIGEN

MDR report key: 3053189 · Received April 11, 2013

Report

Report Number
2124215-2013-04885
Event Type
Malfunction
Date Received
April 11, 2013
Date of Event
March 11, 2013
Report Date
March 11, 2013
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE PRODUCT WAS PERFORMED. THE DEVICE HAD NO TELEMETRY UPON RETURN. ENGINEERING CALCULATIONS CONFIRMED THE LONGEVITY OF THIS DEVICE WAS NOT AS EXPECTED. THE DEVICE CASE WAS OPENED. AN EXTERNAL POWER SUPPLY WAS CONNECTED TO THE DEVICE AND THE ELECTRICAL CURRENT WAS MONITORED. DURING THE MONITORING PROCESS A HIGH CURRENT CONDITION WAS OBSERVED. ELECTRICAL TESTING AND ANALYSIS WERE THEN CONDUCTED, WHICH ISOLATED THE HIGH CURRENT TO AN ANOMALY IN AN OXIDE LAYER WITHIN AN INTEGRATED CIRCUIT. THIS ANOMALY CAUSES A HIGH CURRENT DRAIN, WHICH OVER TIME RESULTED IN REPORTED CLINICAL OBSERVATION(S).

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT NO DEVICE FAILURE DESCRIPTION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
154420 TELIGEN IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND E110

Patients

Seq Age Sex Outcome Treatment
1 72 YR 0185| 1888TC| E110