TELIGEN
Report
- Report Number
- 2124215-2013-04885
- Event Type
- Malfunction
- Date Received
- April 11, 2013
- Date of Event
- March 11, 2013
- Report Date
- March 11, 2013
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWS
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL
- Reporter Occupation
- PHYSICIAN
Narratives
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE PRODUCT WAS PERFORMED. THE DEVICE HAD NO TELEMETRY UPON RETURN. ENGINEERING CALCULATIONS CONFIRMED THE LONGEVITY OF THIS DEVICE WAS NOT AS EXPECTED. THE DEVICE CASE WAS OPENED. AN EXTERNAL POWER SUPPLY WAS CONNECTED TO THE DEVICE AND THE ELECTRICAL CURRENT WAS MONITORED. DURING THE MONITORING PROCESS A HIGH CURRENT CONDITION WAS OBSERVED. ELECTRICAL TESTING AND ANALYSIS WERE THEN CONDUCTED, WHICH ISOLATED THE HIGH CURRENT TO AN ANOMALY IN AN OXIDE LAYER WITHIN AN INTEGRATED CIRCUIT. THIS ANOMALY CAUSES A HIGH CURRENT DRAIN, WHICH OVER TIME RESULTED IN REPORTED CLINICAL OBSERVATION(S).
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT NO DEVICE FAILURE DESCRIPTION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 154420 | TELIGEN | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | GUIDANT CRM CLONMEL IRELAND | E110 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 72 YR | 0185| 1888TC| E110 |