ENDOTAK RELIANCE
Report
- Report Number
- 2124215-2013-04337
- Event Type
- Malfunction
- Date Received
- April 11, 2013
- Date of Event
- March 11, 2013
- Report Date
- May 8, 2013
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- NVY
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE LEAD WAS PERFORMED. RESISTANCE AND PRESSURE TESTS WERE COMPLETED TO ASSESS LEAD ELECTRICAL PERFORMANCE AND INSULATION INTEGRITY. MEASUREMENTS THROUGHOUT THESE TESTS WERE WITHIN NORMAL LIMITS. MICROSCOPIC INSPECTIONS OF THE TERMINAL PIN ASSEMBLY, LEAD BODY, AND ELECTRODE TIP FOUND NO ANOMALIES. LABORATORY TESTING WAS UNABLE TO REPRODUCE THE REPORTED CLINICAL OBSERVATIONS AND DETAILED ANALYSIS DID NOT REVEAL ANY ABNORMALITIES.
THE LEAD WAS REMOVED AND RETURNED FOR ANALYSIS. AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.
--
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT DURING A RECENT IMPLANT PROCEDURE THIS RIGHT VENTRICULAR (RV) LEAD REVEALED A LOSS OF CAPTURE (LOC) DURING PRODUCT TESTING. THE LEAD WAS REMOVED AND REPLACED. NO ADVERSE PATIENT EFFECTS WERE REPORTED. THE PROCEDURE WAS SUCCESSFULLY COMPLETED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 155180 | ENDOTAK RELIANCE | IMPLANTABLE LEAD | NVY | CPI - DEL CARIBE | 0175 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |