FDA Adverse Event Injury Summary report: N

FLEXTEND II

MDR report key: 3053112 · Received April 11, 2013

Report

Report Number
2124215-2013-04653
Event Type
Injury
Date Received
April 11, 2013
Date of Event
March 12, 2013
Report Date
April 15, 2013
Manufacturer
CPI - DEL CARIBE
Product Code
NVN
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE PRODUCT IS EXPECTED TO BE RETURNED FOR ANALYSIS. THIS REPORT WILL BE UPDATED UPON RETURN AND COMPLETION OF ANALYSIS.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE LEAD WAS PERFORMED. RESISTANCE AND PRESSURE TESTS WERE COMPLETED TO ASSESS LEAD ELECTRICAL PERFORMANCE AND INSULATION INTEGRITY. MEASUREMENTS THROUGHOUT THESE TESTS WERE WITHIN NORMAL LIMITS. MICROSCOPIC INSPECTIONS OF THE TERMINAL PIN ASSEMBLY, LEAD BODY, AND ELECTRODE TIP FOUND NO ANOMALIES. LABORATORY TESTING WAS UNABLE TO REPRODUCE THE REPORTED CLINICAL OBSERVATIONS AND DETAILED ANALYSIS DID NOT REVEAL ANY ABNORMALITIES. LABORATORY ANALYSIS DID NOT IDENTIFY ANY LEAD CHARACTERISTICS THAT WOULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED CLINICAL OBSERVATION OF LEAD DISLODGEMENT.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT ONE DAY POST-IMPLANT, THIS RIGHT ATRIAL (RA) LEAD WAS FOUND TO BE DISLODGED. A REVISION PROCEDURE WAS PERFORMED AND THIS LEAD WAS EXPLANTED AND REPLACED WITH A COMPETITIVE LEAD. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Description of Event or Problem · 1

--

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
154602 FLEXTEND II IMPLANTABLE LEAD NVN CPI - DEL CARIBE 4096

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| L| R