INSIGNIA
Report
- Report Number
- 2124215-2013-04338
- Event Type
- Malfunction
- Date Received
- April 11, 2013
- Date of Event
- May 15, 2008
- Report Date
- March 7, 2013
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- NVZ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, DETAILED MECHANICAL AND ELECTRICAL TESTING WAS PERFORMED ON THE DEVICE. THE DEVICE FUNCTIONED NORMALLY THROUGHOUT TESTING. LABORATORY ANALYSIS DETERMINED THAT THIS DEVICE EXPERIENCED NORMAL BATTERY DEPLETION; HOWEVER, THE ESTIMATED LONGEVITY REMAINING VALUE APPEARED TO DECREASE MORE QUICKLY THAN EXPECTED BETWEEN ROUTINE FOLLOW-UPS. FACTORS INFLUENCING THE ESTIMATED LONGEVITY REMAINING CALCULATION INCLUDE PACING RATE, AMPLITUDE, PULSE-WIDTH AND LEAD IMPEDANCE. ANY (EVEN SLIGHT) CHANGES IN THESE FACTORS WILL IMPACT THE BATTERY CONSUMPTION CALCULATION AND THEREFORE THE REMAINING LONGEVITY ESTIMATE. PLEASE NOTE THAT, DESPITE THE DROP IN ESTIMATED LONGEVITY REMAINING, THE ACTUAL BATTERY CONDITION DID NOT CHANGE SIGNIFICANTLY BETWEEN FOLLOW-UPS. IN SUMMARY, IT WAS DETERMINED THAT THIS DEVICE EXPERIENCED NORMAL BATTERY DEPLETION.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT DURING A DEVICE INTERROGATION IT WAS NOTED THAT THE BATTERY LONGEVITY WAS AT THREE AND A HALF YEARS REMAINING. SIX MONTHS AGO THE BATTERY LONGEVITY WAS FIVE YEARS REMAINING. BOSTON SCIENTIFIC TECHNICAL SERVICES (TS) WAS CONSULTED AND PERFORMED A LONGEVITY CALCULATION WHICH SHOWED APPROXIMATELY FOUR YEARS REMAINING LONGEVITY. THE DEVICE WAS EXPLANTED OVER FOUR YEARS LATER FOR REPORTED NORMAL BATTERY DEPLETION. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 154600 | INSIGNIA | IMPLANTABLE PULSE GENERATOR | NVZ | GUIDANT CRM CLONMEL IRELAND | 1290 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 73 YR | 4086| 1290| 4087 |