ALTRUA
Report
- Report Number
- 2124215-2013-04433
- Event Type
- Injury
- Date Received
- April 11, 2013
- Date of Event
- March 12, 2013
- Report Date
- October 8, 2013
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- NVZ
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
ACCORDING TO AVAILABLE INFORMATION, NO RETURN OF PRODUCT IS INTENDED, THEREFORE, BOSTON SCIENTIFIC CANNOT CONFIRM NOR DENY THE REPORTED CLINICAL ALLEGATION. AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE PACING AND SENSING FUNCTIONS WERE TESTED. THE DEVICE OPERATED APPROPRIATELY, ACCORDING TO ITS PERFORMANCE SPECIFICATIONS, WITH NO INTERRUPTIONS IN THERAPY OUTPUT OR PROGRAMMER COMMUNICATION AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED.
UPON COMPLETION OF THE ANALYSIS, THIS EVENT WILL BE UPDATED.
- -
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEVICE WITH NON-BOSTON SCIENTIFIC RIGHT VENTRICULAR LEAD DISPLAYED INCREASED THRESHOLDS AT HIGH OUTPUT SETTINGS. DURING A FOLLOW UP VISIT, INTERROGATION REVEALED THIS DEVICE HAD REACHED END OF LIFE (EOL). THERE WAS CONCERN THAT THE BATTERY DEPLETED MORE RAPIDLY THAN EXPECTED. A REPLACEMENT PROCEDURE WAS PERFORMED. THIS DEVICE WAS REPLACED SUCCESSFULLY. NO RETURN OF PRODUCT IS INTENDED. IN ADDITION, THE RV LEAD WAS SURGICALLY ABANDONED AND REPLACED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
SUBSEQUENTLY, THIS DEVICE WAS RECEIVED FOR ANALYSIS AT THE POST MARKET QUALITY ASSURANCE LABORATORY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 156885 | ALTRUA | IMPLANTABLE PULSE GENERATOR | NVZ | GUIDANT CRM CLONMEL IRELAND | S503 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| L| R |