FDA Adverse Event Injury Summary report: N

ALTRUA

MDR report key: 3053105 · Received April 11, 2013

Report

Report Number
2124215-2013-04433
Event Type
Injury
Date Received
April 11, 2013
Date of Event
March 12, 2013
Report Date
October 8, 2013
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
NVZ
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ACCORDING TO AVAILABLE INFORMATION, NO RETURN OF PRODUCT IS INTENDED, THEREFORE, BOSTON SCIENTIFIC CANNOT CONFIRM NOR DENY THE REPORTED CLINICAL ALLEGATION. AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE PACING AND SENSING FUNCTIONS WERE TESTED. THE DEVICE OPERATED APPROPRIATELY, ACCORDING TO ITS PERFORMANCE SPECIFICATIONS, WITH NO INTERRUPTIONS IN THERAPY OUTPUT OR PROGRAMMER COMMUNICATION AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED.

Additional Manufacturer Narrative · 1

UPON COMPLETION OF THE ANALYSIS, THIS EVENT WILL BE UPDATED.

Description of Event or Problem · 1

- -

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEVICE WITH NON-BOSTON SCIENTIFIC RIGHT VENTRICULAR LEAD DISPLAYED INCREASED THRESHOLDS AT HIGH OUTPUT SETTINGS. DURING A FOLLOW UP VISIT, INTERROGATION REVEALED THIS DEVICE HAD REACHED END OF LIFE (EOL). THERE WAS CONCERN THAT THE BATTERY DEPLETED MORE RAPIDLY THAN EXPECTED. A REPLACEMENT PROCEDURE WAS PERFORMED. THIS DEVICE WAS REPLACED SUCCESSFULLY. NO RETURN OF PRODUCT IS INTENDED. IN ADDITION, THE RV LEAD WAS SURGICALLY ABANDONED AND REPLACED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Description of Event or Problem · 1

SUBSEQUENTLY, THIS DEVICE WAS RECEIVED FOR ANALYSIS AT THE POST MARKET QUALITY ASSURANCE LABORATORY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
156885 ALTRUA IMPLANTABLE PULSE GENERATOR NVZ GUIDANT CRM CLONMEL IRELAND S503

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| L| R