FDA Adverse Event Malfunction Summary report: N

TAXUS¿ LIBERTÉ¿

MDR report key: 3053093 · Received April 11, 2013

Report

Report Number
2134265-2013-02360
Event Type
Malfunction
Date Received
April 11, 2013
Date of Event
September 8, 2012
Report Date
March 14, 2013
Manufacturer
BOSTON SCIENTIFIC - MAPLE GROVE
Product Code
NIQ
PMA / PMN Number
P060008
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IN
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

DEVICE IS COMBINATION PRODUCT. (B)(4). DEVICE EVALUATED BY MFR.: RETURNED PRODUCT CONSISTED OF A TAXUS LIBERTE STENT DELIVERY SYSTEM (SDS) AND NO STENT WITH NO OTHER DEVICES. THERE WAS CONTRAST IN THE INFLATION LUMEN. THERE WERE STENT STRUT IMPRESSIONS PRESENT ON THE SURFACE OF THE BALLOON BETWEEN THE MARKER BANDS, INDICATING THE STENT WAS APPROPRIATELY POSITIONED AND SECURED TO THE BALLOON IN MANUFACTURING. THE BALLOON WAS TIGHTLY FOLDED WHICH IS CONSISTENT OF HAVING NO POSITIVE PRESSURE APPLIED. THERE WAS DISTAL TIP DAMAGE. MAGNIFIED INSPECTION PRESENTED NO DAMAGE OR IRREGULARITIES THAT COULD HAVE CAUSED OR CONTRIBUTED TO THE REPORTED CROSSING DIFFICULTY OR THE STENT DETACHMENT AND TIP DAMAGE. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT AS DEVICE PERFORMANCE WAS LIMITED DUE TO ANATOMICAL/PROCEDURAL FACTORS. (B)(4).

Description of Event or Problem · 1

REPORTABLE BASED ON ANALYSIS COMPLETED ON (B)(4) 2013. IT WAS REPORTED THAT DURING A PERCUTANEOUS CORONARY STENTING TREATMENT PROCEDURE A STENT WAS UNABLE TO CROSS THE LESION, AND A SHAFT KINK OCCURRED. VASCULAR ACCESS WAS OBTAINED VIA THE FEMORAL ARTERY. THE 70% STENOSED <=45 BENT 2.25 X 16MM CONCENTRIC DE NOVO LESION WAS LOCATED IN THE MODERATELY CALCIFIED AND MILDLY TORTUOUS LEFT CIRCUMFLEX ARTERY. THE LESION WAS PREDILATED WITH A NON BSC BALLOON. THE 2.25 X 16MM TAXUS LIBERTE MR STENT DELIVERY SYSTEM WAS ADVANCED, BUT WAS UNABLE TO CROSS THE LESION AND THE SHAFT KINKED. THE PROCEDURE WAS COMPLETED WITH A DIFFERENT DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENT'S STATUS IS STABLE. HOWEVER, DEVICE ANALYSIS REVEALED STENT DISLODGEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
156894 TAXUS¿ LIBERTÉ¿ CORONARY DRUG-ELUTING STENT NIQ BOSTON SCIENTIFIC - MAPLE GROVE H7493894016220 15148481

Patients

Seq Age Sex Outcome Treatment
1 65 YR