FDA Adverse Event Malfunction Summary report: N

COGNIS

MDR report key: 3053091 · Received April 11, 2013

Report

Report Number
2124215-2013-04448
Event Type
Malfunction
Date Received
April 11, 2013
Date of Event
March 13, 2013
Report Date
March 13, 2013
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
NIK
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WV
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, THE DEVICE WAS EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE DEFIBRILLATION, PACING, AND SENSING FUNCTIONS WERE TESTED. THE DEVICE OPERATED APPROPRIATELY WITH NO INTERRUPTIONS IN THERAPY OUTPUT OR PROGRAMMER COMMUNICATION AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WERE ALSO PERFORMED AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED. A REVIEW OF DEVICE MEMORY NOTED THE DEVICE HAD RECORDED ONE OR TWO FAULTS THAT WERE MOST LIKELY CAUSED BY THE USE OF ELECTROCAUTERY.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEVICE DID NOT PASS A PORTION OF THE RETURNED PRODUCTS TESTING, SUGGESTING A POSSIBLE PERFORMANCE ISSUE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
154596 COGNIS IMPLANTABLE CHF PULSE GENERATOR NIK GUIDANT CRM CLONMEL IRELAND N119

Patients

Seq Age Sex Outcome Treatment
1 88 YR 4542| 7122| N119| 4592| 4136