EASYTRAK 2
Report
- Report Number
- 2124215-2013-04297
- Event Type
- Injury
- Date Received
- April 11, 2013
- Date of Event
- March 11, 2013
- Report Date
- April 11, 2013
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- OJX
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
UPON RECEIPT TO THE POST MARKET QUALITY ASSURANCE LABORATORY, ANALYSIS WILL BE PERFORMED AND THIS EVENT WILL BE UPDATED.
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, THIS LEAD WAS CONFIRMED TO BE FRACTURED 41.7 CM MM FROM THE TERMINAL PIN. BASED ON THE ANALYSIS RESULTS, WE SUSPECT THAT FATIGUE OR STRESS IN THE REGION OF THE FRACTURE SITE DUE TO RELATIVE MOTION BETWEEN THE SUTURE SLEEVE AND AN ANATOMICAL FEATURE LED TO THE FRACTURE. AS A LEAD MOVES IN RESPONSE TO NORMAL HEART RHYTHMS AND BLOOD FLOW, EXTENSIVE FLEXING OVER A PERIOD OF TIME MAY CAUSE FATIGUE OR STRESS, WEAKENING THE COIL, ULTIMATELY RESULTING IN A FRACTURE. THIS CAN OCCUR BETWEEN THE SUTURE SLEEVE AND SOME OTHER PART OF THE ANATOMY. FATIGUE FRACTURES IN THE POCKET OR CLAVICULAR/FIRST RIB AREA ARE WELL KNOWN AND DOCUMENTED IN THE INDUSTRY. A COMBINATION OF LEAD DESIGN, IMPLANT TECHNIQUES, AND PATIENT ANATOMY AND ACTIVITY LEVEL CONTRIBUTE TO THESE TYPES OF OCCURRENCES.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT DURING A FOLLOW UP VISIT, THIS LEFT VENTRICULAR LEAD DISPLAYED HIGH OUT OF RANGE IMPEDANCE MEASUREMENTS AND LOSS OF CAPTURE. A LEAD FRACTURE WAS SUSPECTED. A REVISION PROCEDURE WAS PERFORMED. THIS LEAD WAS REMOVED AND REPLACED SUCCESSFULLY. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
- -
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 154585 | EASYTRAK 2 | IMPLANTABLE LEAD | OJX | CPI - DEL CARIBE | 4517 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Hospitalization| L| R |