FINELINE II
Report
- Report Number
- 2124215-2013-04716
- Event Type
- Malfunction
- Date Received
- April 11, 2013
- Date of Event
- March 20, 2013
- Report Date
- April 23, 2013
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- NVN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE LEAD WAS PERFORMED. RESISTANCE AND PRESSURE TESTS WERE COMPLETED TO ASSESS LEAD ELECTRICAL PERFORMANCE AND INSULATION INTEGRITY. MEASUREMENTS THROUGHOUT THESE TESTS WERE WITHIN NORMAL LIMITS. MICROSCOPIC INSPECTIONS OF THE TERMINAL PIN ASSEMBLY AND ELECTRODE TIP FOUND NO ANOMALIES. THERE WAS A VERY SLIGHT BEND/CURVE NOTED IN THE LEAD IN THE NECK REGION. LABORATORY TESTING WAS UNABLE TO REPRODUCE THE REPORTED CLINICAL OBSERVATIONS AND DETAILED ANALYSIS DID NOT REVEAL ANY ABNORMALITIES.
THE PRODUCT IS EXPECTED TO BE RETURNED FOR ANALYSIS. THIS REPORT WILL BE UPDATED UPON RETURN AND COMPLETION OF ANALYSIS.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT RIGHT ATRIAL (RA) LEAD WAS NOT SUCCESSFULLY IMPLANTED DUE TO ABNORMAL IMPEDANCE MEASUREMENTS OF 220 OHMS AND HIGH THRESHOLD MEASUREMENTS. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 156070 | FINELINE II | IMPLANTABLE LEAD | NVN | CPI - DEL CARIBE | 4480 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |