FDA Adverse Event Malfunction Summary report: N

FINELINE II

MDR report key: 3053086 · Received April 11, 2013

Report

Report Number
2124215-2013-04716
Event Type
Malfunction
Date Received
April 11, 2013
Date of Event
March 20, 2013
Report Date
April 23, 2013
Manufacturer
CPI - DEL CARIBE
Product Code
NVN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE LEAD WAS PERFORMED. RESISTANCE AND PRESSURE TESTS WERE COMPLETED TO ASSESS LEAD ELECTRICAL PERFORMANCE AND INSULATION INTEGRITY. MEASUREMENTS THROUGHOUT THESE TESTS WERE WITHIN NORMAL LIMITS. MICROSCOPIC INSPECTIONS OF THE TERMINAL PIN ASSEMBLY AND ELECTRODE TIP FOUND NO ANOMALIES. THERE WAS A VERY SLIGHT BEND/CURVE NOTED IN THE LEAD IN THE NECK REGION. LABORATORY TESTING WAS UNABLE TO REPRODUCE THE REPORTED CLINICAL OBSERVATIONS AND DETAILED ANALYSIS DID NOT REVEAL ANY ABNORMALITIES.

Additional Manufacturer Narrative · 1

THE PRODUCT IS EXPECTED TO BE RETURNED FOR ANALYSIS. THIS REPORT WILL BE UPDATED UPON RETURN AND COMPLETION OF ANALYSIS.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT RIGHT ATRIAL (RA) LEAD WAS NOT SUCCESSFULLY IMPLANTED DUE TO ABNORMAL IMPEDANCE MEASUREMENTS OF 220 OHMS AND HIGH THRESHOLD MEASUREMENTS. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
156070 FINELINE II IMPLANTABLE LEAD NVN CPI - DEL CARIBE 4480

Patients

Seq Age Sex Outcome Treatment
1